Eon 300EonTM 300 - Technical Specifications

Clinical Chemistry

The next generation of advanced chemistry systems

PRODUCTIVITY
Throughput 300 photometric tests per hour
120 ISE tests per hour
Up to 500 tests per hour combined photometric and ISE
Time to first result Photometric: 2 - 4 min plus incubation time
ISE : 1 minute
Sample addition Continuous loading into 69 positions
Sample types Serum, Plasma, and Whole Blood (HbA1c)
Reagent storage On board @ 12°-14°C below ambient (71 positions)
Store multiple bottles of the same reagent (same Lot number only)
Reaction cuvette On board 8 step cuvette laundry system (80 positions)
Water consumption 4 L/h, no requirement for direct supply connection
Identification Integrated sample and reagent barcode reader
Measurement System
Photometer Halogen lamp (2,000hrs); 9 λs(340-700nm);mono and
bichromatic measurement
Optical linearity to 3 Absorbance Units (1cm)
Temperature controlled (37° ±0.1°C) optical plastic cuvette
Minimum reaction volume 200µL
Analytical modes Kinetic Rate; endpoint; 2-Point rate (fixed time); differential sample blank;
differential 2- reagents; and direct ISE
Sample & Reagent management
Sample and Reagent dispensing Dual probe with pre heater, liquid level and shock sensors
Sample containers Primary tubes (12-16mm diam / 75-100 mm height) (200μL dead volume) 3 mL sample cups (50μL dead volume)
Sample Volume Minimum 1μL: Maximum 100μL
Dilution Automatic pre and post analytical (rerun) dilution of sample
Reagent Containers Bar coded system reagent bottles
Reagent consumption Up to 3 reagents per test
Total reagent volume typically no more than 200μL per test
Calibration and Quality Control
Capacity Up to 69 positions (controls and calibrators)
Calibration modes Single point; fixed factor; and non linear
Automatic standard curve dilution, up to 8 points.
QC monitoring Westgard rules
Levey –Jennings Plots
data management
Operating System Windows XP Pro/embedded/ External PC
Control Unit Colour screen / keyboard
Result Storage QC (5,000 tests) and Patient (15,000 tests) archive
System Interface LIS connection; RS232 bidirectional serial interface
Result Output Graphical plot of kinetic & fixed time reactions; and calibration curves
Physical Characteristics / Requirements
Dimensions 110L x 55H x 70D (cm)
Weight 70 Kg
Electrical Requirements Line Voltage 100-125V or 200-240V (±10%)
Line Frequency 50 Hz - 60 Hz
Regulatory Approvals CE marking -IVD

 